Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 78: CAL ADJ /5520 DATE: 19-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 53 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 78 Service Manual (PN 926923, March 1993, Errata No. 2 9/94). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- W 1.002 HEAD CALIBRATION ACCESS PROCEDURE 1.003 DISP Select the V DC function. 1.003 DISP 1.003 DISP Press the push button five times to select AUTO 4V Range 1.004 DISP Remove the four phillips-head screws from the bottom of 1.004 DISP the case. Turn the instrument face-up, and grasp the top 1.004 DISP cover at both sides. Pull the top cover from the UUT. 1.005 HEAD EQUIPMENT SETUP 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 80%. 1.006 DISP Allow the UUT to stabilize. 1.007 HEAD CALIBRATION ADJUSTMENT 1.008 DISP Connect the 5520A and the UUT as follows. 1.008 DISP [32] 5520A NORMAL HI to UUT V Ohm Diode 1.008 DISP [32] 5520A NORMAL LO to UUT COM 1.009 5520 3.5V S 2W 1.010 DISP Adjust R26 for a UUT a reading of 3.500+/-0.001V. 1.011 END